Clinical evaluation of a RT-LAMP SARS-CoV-2 test for the Point-Of-Care, rapid, low-cost, integrating sample solid phase extraction and on which reagents are lyophilized - Institut Pasteur
Pré-Publication, Document De Travail Année : 2021

Clinical evaluation of a RT-LAMP SARS-CoV-2 test for the Point-Of-Care, rapid, low-cost, integrating sample solid phase extraction and on which reagents are lyophilized

A. Vilquin
  • Fonction : Auteur

Résumé

Objectives Determine the sensitivity and specificity of a Point-Of-Care test (‘COVIDISC’) for SARS-COV2. The novelty of the test is to integrate, on the same (low-cost) compact plastic/paper device, solid phase RNA extraction and RT-LAMP amplification, all reagents being freeze-dried on it. Method Retrospective study with a cohort of 99 patients characterized by real-time RT-PCR. The 37 positive naso-pharyngeal samples cover a broad range of viral loads (from 5 gc /µL to 2 10 6 gc/ µL of sample). Results The COVIDISC found 36 positives (out of 37 by IP4 RT-PCR protocols) and 63 negatives (out of 62 by RT-PCR). Conclusion The sensitivity of the COVIDISC, found in this 99-patient retrospective study, is 97% and the specificity 100%.
Fichier principal
Vignette du fichier
2021.10.03.21264480v1.full.pdf (2.39 Mo) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Licence
Copyright (Tous droits réservés)

Dates et versions

pasteur-04670582 , version 1 (12-08-2024)

Licence

Copyright (Tous droits réservés)

Identifiants

Citer

E. Coz, P. Garneret, E. Martin, D. Do Nascimento, A. Vilquin, et al.. Clinical evaluation of a RT-LAMP SARS-CoV-2 test for the Point-Of-Care, rapid, low-cost, integrating sample solid phase extraction and on which reagents are lyophilized. 2024. ⟨pasteur-04670582⟩
105 Consultations
12 Téléchargements

Altmetric

Partager

More